| PARAMETER |
SPECIFICATION |
| Input Voltage |
230V ±10% AC, 50Hz, single phase |
| Maximum Output Voltage & Power * |
| XRM10 (10kW) |
XRM15 (15kW) |
| 40kV at 250mA |
50kV at 300mA |
| 100kV at 100mA |
120kV at 125mA |
|
| HF Generator Frequency |
80kHz |
| Protections |
Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Rotor Fault, Inverter Fault. |
| Application |
Intensive care units and Radiology department |
| kV Range |
40kV to 120kV (80 steps) |
| mA Range |
25mA to 250mA / 300mA (Depends on the Power)(8 Steps) |
| Modes |
APR & Manual |
| mAs |
0.1 to 200mAs in 31steps |
| Time |
10ms to 5sec (Depends on the Capacitor Bank) |
| User Interface |
Keyboard or Touch screen tablet with microcontroller-based control |
| Housing |
Monoblock |
|
HF Inverter |
|
| Contains high voltage transformer, X-ray tube & filament transformers floating on high voltage |
| Contains inverter, control & filament cards, rotor circuit, transformers and other interlocks |
|
| Size (L x W x H) |
Monoblock |
|
HF Inverter |
|
| 311mm x 156mm x 267mm |
| 491mm x 336mm x 700mm |
|
| Weight |
Monoblock |
|
HF Inverter |
|
|