| PARAMETER |
SPECIFICATION |
| Gun Injection Power Supply |
| Pulse height |
Variable from 0kV to -35kV peak(max) |
| Maximum Current |
3A peak |
| Pulse shape |
Rectangular |
| Pulse width (measured at the flat top) |
Adjustable from 4μS to 12μS |
| Ruse time |
2.5μS |
| Flat top variation (peak to peak) |
Flatness within 1% with initial overshoot less than 5% of the set value |
| Pulse Repetition Frequency (RPF) |
Variable from 1 to 250 pps |
| Maximum reverse voltage |
Less than 1.5kV |
| Maximum average power |
360W |
| Cathode Heater Power Supply |
| Output Voltage |
10V A.C. max |
| Output Current |
10V A.C. max |
| Regulation |
Poor regulation |
| Output Reference |
Floating at -40kV (output of the gun injection power supply) |
| Output isolation |
Minimum -40kV w.r.t. earth under pulsed condition. (Pulsed width: 12μS, PRF:250Hz) |
| Trip |
10Amp |
| Grid Power Supply (-1kV DC Section) |
| Output Voltage |
-1kV DC settable in steps of 100V; the Grid is at negative potential with respect to the cathode |
| Output Current |
200mA DC max |
| Pulse shape |
Rectangular |
| Type of control |
Voltage regulated with load regulation <_ ±0.1% |
| Output reference |
Floating at -40kV (output of the pulsed gun injection power supply section) |
| Output isolation |
Minimum -45kV w.r.t earth under DC condition. |
| Grid Power Supply (1kV PULSED Section) |
| Output Voltage (Pulsed) |
Pulsed amplitude settable from 1V to 1kV DC precisely variable in continuous manner. This pulse overrides the -1kV DC voltage at the grid |
| Pulse Shape |
Rectangular |
| Pulse width (measured at the flat top) |
Adjustable from 4μS to 10μS |
| Rise time |
0.5μS |
| Flat top variation (peak to peak) |
Flatness within 1% with initial overshoot less than 5% of the set value. |
| Output Current |
10mA max. |